Sanitary Electromagnetic Flowmeter
Introduction
The electromagnetic flowmeter uses Faraday’s Law of
electromagnetic induction to measure the process flow.
The device consists of two units: a detector, through
which
the fluid to be measured flows and in which
low-level signals proportional to flow rates are
obtained; and a converter, which supplies excitation
current to the
detector, and amplifies the signals from
the detector and then processes and converts the
signals into the 4–20 mA dc current signal.
The LF232 offers various outputs such as totalization
output and multi-range selection outputs. With it
patented Noise-Sentry original noise-suppression
circuit and advanced algorithms. The LF232 is highly
resistant to noise and provides a stable output even
when measuring fluids contain slurries. The
LF232 can
be used with a separately mounted detector such as the
LF150 or LF470. The LF232 can also be used with
conventional models of detectors.
The AF900 hand-held terminal (HART*
1
communicator) can be used to communicate with the
flowmeter from a remote place. See the
Communications signal specification for details about
HART protocol.
*1: HART protocol (Highway Addressable Remote
Transducer) is a communication protocol for industrial
sensors recommended by the HCF
(HART
Communication Foundation).
Specifications
Overall Specifications
Measurement range in terms of flow velocity:
Conbination LF150, LF470
0 –1.0 ft/s to 0 – 32.8 ft/s (0 – 0.3 m/s to 0 –10 m/s)
Fluid conductivity: 5µS/cm or more
Ambient temperature: -4 to 140 °F (-20 to 60 °C)
(Storage temperature: -13 to 149 °F (-25 to 65 °C))
Structure: IP67 and NEMA 4X Watertight
Figure 1. LF232 Converter
Accuracy: the accuracy is dependent on the type of
detector combined with the LF232. See the following
tables for accuracy when combined with specified
detectors.
Detector combined
LF470 (Meter size 1/10", 1/6, 1/4"
(2.5 to 6 mm))
Accuracy Flow rate as a
percent of range
0.3–1.0m/s 1.0–10m/s
050 ±0.8FS
±0.4FS
50100 ±0.8FS ±0.8 of rate
Detector combined
LF150 (Meter size 28” to 120” (700 to 3000 mm))
Accuracy: ±0.5 % of Rate*
* This pulse output error result is established under
standard operating conditions at Toshiba’s flow
calibration facility, Fuchu Japan.
* Individual meter’s measurement error may vary up
to ± 0.8 % of Rate at 3.28 ft/s (1.0 m/s) or more and
± 0.4% of Rate ±0.157 inch/s (4mm/s) at 3.28 ft/s
(1.0 m/s) or less.
* Current output: plus ± 8µA (0.05% of span).
* Refer to individual calibration data for each
individual meter’s measurement error.
LF232
Field Intelligent Device Series
Electromagnetic Flowmeter Converter
Vue de la page 0
1 2 3 4 5 6 7 8

Résumé du contenu

Page 1 - Specifications

TIC-LF232A Field Intelligent Device Series Sanitary Electromagnetic Flowmeter Introduction The electromagnetic flowmeter uses Faraday’s Law of el

Page 2

TIC-LF232A 2 Power supply: One of the following can be selected ・100 to 240 Vac (Allowable voltage 80 to 264 Vac, 50/60Hz)*1 ・24 Vdc (Allowable

Page 3 - Mag-Prover – Toshiba’s

TIC-LF232A 3Communications signal: A digital communications (HART protocol) signal is superimposed on 4 to 20 mA dc analog output signal. Load resis

Page 4 - Installation

TIC-LF232A 4Installation  Dimensions 8.74 (222) 2.91 (74) 9.33 (237) 4-φ12 hole 6.10 (155) 14.57(370) 15.75 (400) 2.99 (76) 2.99 (76

Page 5 - TIC-LF232A

TIC-LF232A 5  External Connections Figure 3. External Wiring Diagram Power supply Current output (

Page 6 - Ordering Information

TIC-LF232A 6  Wiring Precautions (1)Be sure to use thick walled steel conduit (22mm) for signal and excitation cable wiring between the detector

Page 7

TIC-LF232A 7Table 1. Specification Code for converters (LF232) Model Specification Code 1 2 3 4 5 6 7 8 9 10 11 12 13 14Contents L F 2 3 2

Page 8 - TIC-LF232A

TIC-LF232 TIC-LF232A Misuse of this product can result in damages to property or human injury. Read related manuals carefully before using this pr